Читать книгу Essentials of MRI Safety - Donald W. McRobbie - Страница 66

MYTHBUSTER:

Оглавление

The internal magnetic field or degree of magnetization of a ferromagnetic object is not determined by its magnetic susceptibility but by its demagnetization factors and saturation status.