Читать книгу Materials for Biomedical Engineering - Mohamed N. Rahaman - Страница 125
Example 4.1
ОглавлениеTensile testing of a metal alloy using a cylindrical dog‐bone specimen of gage length 60.00 mm and diameter 12.8 mm yielded the following data:
Force (N) | Length (mm) |
---|---|
0 | 60.00 |
15 442 | 60.06 |
30 883 | 60.12 |
46 325 | 60.18 |
61 766 | 60.24 |
77 208 | 60.30 |
92 649 | 60.36 |
105 517 | 60.42 |
107 834 | 60.48 |
109 506 | 60.54 |
110 150 | 60.60 |
110 664 | 60.66 |
108 992 | 60.72 |
106 547 | 60.78 |
Determine (a) the Young’s modulus, (b) the elastic limit, and (c) the yield strength of the metal alloy.