Читать книгу Process Gas Chromatographs - Tony Waters - Страница 70

A more realistic explanation

Оглавление

We need a better explanation. Let's start from the obvious truism that when the sample molecules are in the column, they must always be in the gas phase or the liquid phase. There's nowhere else for them to be.

And we know that the liquid phase doesn't move.

Therefore:

 When the sample molecules are dissolved in the liquid phase, they are held in place and are not moving along the column.

 When the sample molecules are in the gas phase, they move along the column at the same speed as the carrier gas.

Note that we refer to movement along the column. Molecules are always moving randomly in every direction, and this contributes to peak broadening, but random motion contributes nothing to positive movement along the column.

There are only two speeds along a chromatograph column; stop or go!

This is true for every peak. Every injected molecule must spend enough time in the gas phase to transit through the column. Therefore, each molecule travels at the same speed and spends the same time traveling as the other molecules do. It's not true to say that different molecules travel at different speeds.

Separation is not caused by motion at all. It's caused by stopping; the time that different molecules stay motionless in the liquid phase. The less soluble molecules don't stop for long and move quickly through the column, while the more soluble ones hang around in the liquid, slowing them down. It's that simple.

Notwithstanding the stop‐go mechanism, one might successfully argue that the peaks really are travelling at different average speeds through the column. Yes, of course, this is the overall result. But talking about the average speed of a peak obscures what is really happening in the column. Instead, chromatographers may refer to the migration rate of the peak along the column. The word “migration” infers a gradual movement by alternately stopping and going, to remind us what is really going on in there.

It should now be clear why the propane peak in Figure 2.9 is exactly in the middle of the column. We assumed that 50 % of the propane molecules dissolve in the liquid phase and 50 % remain in the gas phase. Remember that those molecules are rapidly moving between the two phases. So the average propane molecule spends 50 % of its time in the gas phase moving at carrier speed and 50 % of its time in the liquid phase going nowhere. Therefore, when the carrier gas that was present during sample injection reaches the end of the column, the propane peak is exactly half‐way.

Figure 3.1 applies this logic to two additional peaks, one more soluble and one less soluble than propane. The 1‐butene peak is more soluble, so we would expect it to spend more time in the liquid phase and elute from the column later than the propane. And so it does. For instance, if the solubility of 1‐butene is 75 %, only 25 % of the molecules move each time the carrier gas moves, and the peak stays in the column twice as long as the propane peak does. The carbon dioxide peak is less soluble: If its solubility is 25 %, then 75 % of its molecules move with the carrier gas.

Drawing peaks above a column is a common way to show the position of component molecules within the column at a certain instant of time. Here, an air peak (which doesn't dissolve in the liquid phase) has moved along with the carrier gas and has just arrived at the column end. At that instant, the three colored peak drawings indicate the position within the column of the other component molecules, predictable from their solubility in the liquid phase.

Figure 3.1 Effect of Component Solubility.

This is the true cause of separation. When in the gas phase, all sample molecules move along the column at the same speed as the carrier gas. But when in the liquid phase, the molecules stop moving and the more soluble ones stop longer than the less soluble ones do.

Figure 3.1 illustrates the net effect of solubility difference. It shows the location of four components peaks at the exact moment the air peak reaches the end of the column. A small white‐and‐blue equilibrium diagram indicates the solubility of each peak.

Figure 3.1 assumes that:

 The stationary phase is a liquid, and the air peak is not soluble at all. The air peak therefore moves with the carrier gas, and its retention time is a good indicator of average carrier gas velocity.

 The carbon dioxide (CO2) solubility is 25 %, so an average CO2 molecule spends 75 % of the time traveling and only 25 % of the time stopped. Therefore, the CO2 peak moves 75 % of the distance that the carrier gas moves.

 The propane solubility is 50 %, so an average propane molecule spends half the time traveling and half the time stopped. Therefore, the propane peak moves 50 % of the distance that the carrier gas moves.

 The 1‐butene solubility is 75 %, so an average molecule spends only 25 % of the time traveling and 75 % of the time stopped. It follows that the 1‐butene peak moves only 25 % of the distance that the carrier gas moves.

These convenient values for the component solubilities are just simple examples assumed for discussion purposes, but every substance has a real solubility in a given liquid that depends only on the temperature and pressure.

Figure 3.1 shows the location of each peak when the air peak has reached the end of the column and is about to flow into the detector. At that moment, the four peaks in our example are equally spaced along the column. Similar spacing is a common occurrence in real columns, but a very strange thing then happens to the chromatogram. The exercise that follows challenges you to discover what comes next.

Process Gas Chromatographs

Подняться наверх