Читать книгу Magnetic Resonance Microscopy - Группа авторов - Страница 58

3.4.4 Neonatal Intensive Care Unit (NICU)

Оглавление

A scanner situated in an NICU or capable of being brought to the NICU (beside incubator/isolette) could have many uses. Here we focus on one: HIE to provide a general picture of the uses and issues of MR in a NICU.

HIE and resulting hypoxic-ischemic injury (HII) is the leading cause of neonatal mortality and morbidity with an occurrence of 1 to 8 out of 1000 live births in the United States and up to 26 out of 1000 births in developing countries [74,75]. In survivors, HIE is a predictor of permanent neurodevelopmental disability. The significance of early diagnosis and intervention in HIE is profound, particularly when considering the long-term consequences in remaining years of life.

Once HIE is identified, immediate supportive care is applied to encourage brain perfusion, including intubation and mechanical ventilation for more severely affected neonates, titration of intravenous fluids, and blood pressure support [76]. Continuous electroencephalogram (EEG) monitoring is used to monitor background brain electrical activity and seizures. Expedient intervention with hypothermia neural rescue therapy within the first few hours of life is vital, improving survival rates and reducing future disabilities [77]. Thus, moderate to severe cases undergo neuroprotective interventions with therapeutic hypothermia (also known as hypothermia neural rescue therapy or “cooling”) where a cooling blanket and temperature monitor maintain a body temperature decrease of ~3°C for 72 h. Successful treatment of neonatal brain injury could eliminate many decades of disease burden and healthcare needs.

As the gold standard for brain imaging, MRI is the best imaging technique for diagnosing and staging HIE, particularly with diffusion-weighted contrast [78]. CT is less suitable due to its limited soft tissue contrast and the risks of exposing neonates to ionizing radiation. Although the preferred imaging method, MRI currently requires transport of the fragile patients to a radiology suite. This introduces safety concerns, since it requires a full hand-off of patient care responsibilities and exposes the neonate to stress from transport and acoustic noise. In particular, HIE patients are often not stable enough to transfer to the scanner for several days after birth. Although safety is the number one concern, we must also consider the burden on the hospital workflow. Transfer of critical care patients requires tremendous teamwork and coordination from multiple sources. For example, a neonatologist, nurse, and respiratory therapist must accompany the infant to the scanner and remain with the patient for the duration of the study. The time required for the scan and transfer to/from the NICU is around three hours and NICU staffing coverage must be implemented during their absence. Currently, bedside cranial imaging is available with ultrasound, but provides low sensitivity for early abnormalities associated with HIE [78].

Siting a small footprint MRI in the NICU would provide tangible benefits and a bedside (portable) MRI could offer additional benefits. Figure 3.5 illustrates some existing MRI systems and components that have been developed or adapted for MRI in the NICU. The benefits of bedside MRI accrue since even the trip from one side of the NICU to the other is difficult due to the removal/replacement of leads, tubes, ventilation, and temperature maintenance equipment. Furthermore, intravenous (IV) pumps are not MR compatible so extension tubing must be placed, which can interrupt the administration of critical medications (e.g. dopamine). Even if “MR-safe,” noncompatible or metallic equipment can cause susceptibility artifacts, which are more severe at higher field strengths [79]. All of these issues point to an extremely compact magnetic footprint magnet; likely a mid- to low-field magnet design with minimal fringe field.


Figure 3.5 MR devices designed to support imaging in neonatal intensive care unit (NICU). (a) LMT MRI compatible incubator (LMT Medical Systems GmbH). (b) GE Healthcare investigational 3-T neonatal MR system installed at Sheffield Teaching Hospital in a dedicated NICU scan room (EMAP Publishing Limited). (c) The Hyperfine 64-mT portable MRI scanner. (d) Aspect Imaging’s Embrace neonatal scanner, a dedicated 1-T permanent magnet scanner designed for installation in the NICU without a shielded room. (e) Modified 1.5-T GE orthopedic scanner adapted for the NICU at Cincinnati Children’s Hospital.

Magnetic Resonance Microscopy

Подняться наверх