Читать книгу Fundamentals of Analytical Toxicology - Robin Whelpton - Страница 78
Box 2.5 Guidance on freezer storage of samples
ОглавлениеDo not freeze whole blood if plasma or serum is to be analyzed
Ensure that labelling is waterproof
Ensure tubes are tightly sealed and well filled, but do not overfill tubes, especially glass tubes
Do not keep too long to minimize freeze-drying effects
Keep a record of freezer contents and freezer cleaning
Keep a continuous record of freezer temperature via an electronic device
Use a spark-proof freezer if flammable materials may be stored
Fit an alarm in case of freezer failure
Implement a defined sample disposal policy in collaboration with stakeholders